| Insulation
Resistance |
| Test
Voltage | 250V/500V/1000V |
Measuring
Ranges (Mid-scale Value) |
| 0~100MΩ/ |
0~200MΩ/ |
0~400MΩ |
| (1MΩ) | (2MΩ) | (4MΩ) |
|
Output
Voltage on open circuit | Rated
test voltage ±10% |
| Rated
Current | 0.8mA
DC approx. |
Output
Short Circuit Current | 2mA
DC approx. |
Accuracy
Guaranteed Ranges | 0.05~10MΩ/0.1~20MΩ/0.2~40MΩ |
| Accuracy | |±5%
of indicated value |
| Continuity Test |
Measuring
Ranges (Mid-scale Value) | 50Ω
(20Ω) |
Output
Voltage on open circuit | 0.6V
DC approx. |
Output
Short Circuit Current | 300mA
approx. |
| Accuracy | |±3%
of scale length |
| General |
| Withstand
Voltage | 2200V AC
for 1 minute |
| Power
Source | R6P (AA) (1.5V)
× 8 |
| Dimensions | 143(L)
× 93(W) × 63(D)mm |
| Weight | 500g
approx. |
| Accessories |
7025 (Test
leads)
fuse 250V/1A
9038 (Carrying Case) R6P (AA)
× 8 Instruction Manual |