KYORITSU
HOME

PRODUCTS
   • Multimeters
   • Voltage Testers
   • Clamp Meters
   • Loggers
   • Insulation Testers
   • Earth Testers
   • LOOP/PFC/PSC
     Testers
   • RCD Testers
   • Multi Function Testers
   • Portable Appliance
     Testers
   • Power Meters
   • Sensors, Multi-Trans,
     Clamp Adaptors
   • Others
   • Accessories
   • Discontinued Products

 
Leakage Clamp Meters

KEW 2413F
 
Selection Guide Instruction Manual
 

Extra wide transformer jaws are best suited for clamping on all three or four wires (3phases) together for leakage current measurement.
Frequency filter switch to eliminate the effect of harmonics.
Output terminal.

Ø68 MAX AC1000 Resolution 0.1mA
OUT PUT PEAK 10/100ms Filter
2413F
CE
 
AC A
(50/60Hz)
200mA/2/20/200A/1000A
|±1.5%rdg±2dgt (200mA/2/20A)
|±2%rdg±2dgt (200A 0 - 500A)
|±5.5%rdg (501 - 1000A)
AC A
(WIDE)
200mA/2/20/200/1000A
|±1%rdg±2dgt [50/60Hz],
|±3%rdg±2dgt [40Hz - 1kHz] (200mA/2/20A)
|±1.5%rdg±2dgt [50/60Hz],
|±3.5%rdg±2dgt [40Hz - 1kHz] (200A/0 - 500A)
|±5%rdg [50/60Hz],
|±10%rdg [40Hz - 1kHz] (501 - 1000A)
Conductor Size Ø68mm max.
Effect of External
Stray Magnetic Field
Ø15mm 100A
10mA AC max.
Frequency Response 40Hz - 1kHz
Output
Waveform : AC200mV against the maximum value of each range (1000A range is 100mV)
Recorder : DC200mV against the maximum value of each range (1000A range is 100mV)
Withstand Voltage 3700V AC for 1 minute
Applicable Standard IEC 61010-1 CAT III 300V
IEC 61010-2-032
Power Source 6F22 (9V) × 1 *Continuous measuring time: Approx. 60 hours
Dimensions 250(L) × 130(W) × 50(D)mm
Weight 570g approx.
Accessories 9094 (Carrying Case)
6F22 × 1
Instruction Manual
Optional 7073 (2WAY Output Cord)

PageTop